- chemical definition
FT mid-IR Spectrometer
InfraLUM FT-08

Our new FT mid-IR spectrometer InfraLUM FT-08 has been designed basing on our extensive R&D and expertise in the IR spectroscopy (infrared spectroscopy). This multipurpose and easy-to-operate instrument provides a well suited option for various analytical laboratories. The high-quality misalignment-proof optics, high-sensitivity DLATGS detector, and a wide variety of accessories ensure the users’ absolute confidence in the high accuracy of the measurement data and make it an ideal tool for routine measurements at low inventory investment and running costs.
Measurement procedure
The instrument is operated by SpectraLUM software run under Windows environment, which provides simple control and allows even unskilled personnel to easily master the instrument operation.
All major functions, such as peak identification, derivative quantification, quantitative calibrations creation, etc., can easily be performed with SpectraLUM software. Library search module allows to plug spectra libraries and to create own user’s libraries.
High signal-to-noise ratio
Hermetically sealed optical
compartment with the automated
moisture monitoring system
Enlarged cell compartment
PIKE Technologies accessories
Broad range of spectra database
FEATURES AND BENEFITS
- Patented high-quality misalignment-proof Double Cat’s Eye interferometer
- Hermetically sealed optical compartment with the automated moisture monitoring system
- High-sensitivity DLATGS detector
- Advanced radiation source with extended service life
- Built-in smart total self-diagnostics system
- Innovative ergonomic design of the spectrometer housing that provides best convenience for the operator
- Large sample compartment 230 X 240 X 180mm
- Wide variety of accessories, include complete line of compatible Specac and PIKE Technologies optical FTIR accessories
- Intuitive comprehensible software with integrated calibration module & library search module
- Automatic validation system of the spectrometer
- Remote diagnostics of the spectrometer
Technical specifications
Spectral range | 8000-350 cm-1 (KBr optics)8000-500 cm-1 (ZnSe optics) |
Wavenumber repeatability | 0.01 cm-1 |
Variable resolution | 64, 32, 16, 8, 4, 2, 1, 0.5 cm-1 |
RMS signal-to-noise ratio | > 60000 (4 cm-1, 1 min) |
Stability of the 100% line, % | 0.2 |
Scanning speed | 0.8 s/scan at 16 cm-1 resolution |
Detector | High resolution DLATGS |
Beamsplitter | Ge-coated KBr (or ZnSe) in a hermetically sealed dessicated housing |
Wavenumber accuracy | ±0.05 cm-1 |
Interfereometer | Vibration- and misalignment-proof patentedDouble Cat’s Eye Interferometer |
Radiation source | High-intensity long-life stabilized ceramic IR radiation source |
Power consumption | 65 W |
Weight | 32 kg |
Size | 580x550x340 mm |