Electronic Personal Dosimeter NRF3 Series / NRF40

NRF30
Radiation MeasuredGamma (X) rays
DetectorSilicon semiconductor
Measurement Range1μSv to 9.999Sv (0.1mrem to 999.9rem)
1μSv/h to 9.999Sv/h
(0.1mrem/h to 999.9rem/h) for dose measurement
Accuracy≤ ± 10 %
(0.1mSv to 9.999Sv (10mrem to 999.9rem), 137Cs)
Energy Range30keV to 6MeV
Energy Response≤ ± 20% (50keV to 6MeV, 137Cs reference)
Angular Response≤ ± 20%
(Up to ±60degree vertical and horizontal, 137Cs)
≤ ± 50%
(Up to ±60degree vertical and horizontal, 241Am)
≤ ± 30%
(All around horizontal,137Cs, free air)
Dose Rate Linearity≤ ± 10% (0.1mSv/h to 9.999Sv/h
(10mrem/h to 999.9rem/h),137Cs)
DisplayLCD (with backlight)
Alarm Volume85 to 100dB (at 30cm)
Communication MethodInfrared communication
Operating Temperature-20℃ to 50℃ (-4°F to 122°F)
Operating Humidity35% to 95%RH (Non-condensing)
Power SupplyPrimary Battery (CR123A) x 1
Battery Life2880hours (without alarm/ communication)
Dimensions60(W) x 33(D) x 78(H)mm
(2.4(W) x 1.3(D) x 3.1(H) in)
Weight100g approx. (0.22lb approx.)
Top of Page

NRF31

NRF31
Radiation MeasuredGamma (X) rays and Neutron
DetectorSilicon semiconductor
Measurement RangeGamma(X) rays : ≤ ± 10%
(0.1mSv to 9.999Sv (10mrem to 999.9rem),137Cs)
Neutron : ≤ ± 15%
(0.5mSv to 9.999Sv (50mrem to 999.9rem), 252Cf)
AccuracyGamma(X) rays : ≤ ± 10%
(0.1mSv to 9.999Sv (10mrem to 999.9rem),137Cs)
Neutron : ≤ ± 15%
(0.5mSv to 9.999Sv (50mrem to 999.9rem), 252Cf)
Energy RangeGamma(X) rays : 30keV to 6MeV
Neutron : 0.025eV to 15MeV
Energy ResponseGamma(X) rays : ≤ ± 20% (50keV to 6MeV, 137Cs reference)
Neutron : ≤ ± 50% (250keV to 4.5MeV, 241Am-Be reference)
Angular ResponseGamma(X) rays :
≤ ± 20% (Up to ±60degree vertical and horizontal, 137Cs)
≤ ± 50% (Up to ±60degree vertical and horizontal, 241Am)
≤ ± 30% (All around horizontal, 137Cs, free air)
Neutron :
≤± 30% (Up to ±75degree vertical and horizontal, 241Am-Be)
Dose Rate LinearityGamma(X) rays : ≤ ±10%
(0.1mSv/h to 9.999Sv/h (10mrem/h to 999.9rem/h), 137Cs)
Neutron : ≤ ± 20%
(0.5mSv/h to 9.999Sv/h (50mrem/h to 999.9rem/h), 252Cf)
DisplayLCD (with backlight)
Alarm Volume85 to 100dB (at 30cm)
Communication MethodInfrared communication
Operating Temperature-20℃ to 50℃ (-4°F to 122°F)
Operating Humidity35% to 95%RH (Non-condensing)
Power SupplyPrimary Battery (CR123A) x 1
Battery Life2880hours (without alarm/ communication)
Dimensions60(W) x 33(D) x 78(H)mm
(2.4(W) x 1.3(D) x 3.1(H) in)
Weight120g approx. (0.26lb approx.)
Top of Page

NRF34

NRF34
Radiation MeasuredGamma (X) rays and Beta rays
DetectorSilicon semiconductor
Measurement Range1μSv to 9.999Sv
(0.1mrem to 999.9rem)
1μSv/h to 9.999Sv/h
(0.1mrem/h to 999.9rem/h) for dose measurement
AccuracyGamma(X) rays : ≤ ± 10%
(0.1mSv to 9.999Sv (10mrem to 999.9rem), 137Cs)
Beta rays : ≤ ± 15%
(0.1mSv to 9.999Sv (10mrem to 999.9rem), 90Sr/90Y)
Energy RangeGamma(X) rays : 30keV to 6MeV
Beta rays : 0.2MeV to 0.8MeV (Mean energy)
Energy ResponseGamma(X) rays : ≤ ± 20%
(50keV to 6MeV, 137Cs reference)
Beta rays : ≤ ± 30%
(0.2MeV to 0.8MeV (Mean energy), 90Sr/90Y reference)
Angular ResponseGamma rays :
≤ ± 20% (Up to ±60degree vertical and horizontal, 137Cs)
≤ ± 50% (Up to ±60degree vertical and horizontal, 241Am)
≤ ± 30% (All around horizontal, 137Cs, free air)
Beta rays :
≤ ± 30% (Up to ±60degree vertical and horizontal, 90Sr/90Y)
Dose Rate LinearityGamma(X) rays : ≤ ± 10%
(0.1mSv/h to 9.999Sv/h (10mrem/h to 999.9rem/h), 137Cs)
Beta rays : ≤ ± 20%
(0.1mSv/h to 9.999Sv/h (10mrem/h to 999.9rem/h), 90Sr/90Y
DisplayLCD (with backlight)
Alarm Volume85 to 100dB (at 30cm)
Communication MethodInfrared communication
Operating Temperature-20℃ to 50℃ (-4°F to 122°F)
Operating Humidity35% to 95%RH (Non-condensing)
Power SupplyPrimary Battery (CR123A) x 1
Battery Life2880hours (without alarm/ communication)
Dimensions60(W) x 33(D) x 78(H)mm
(2.4(W) x 1.3(D) x 3.1(H) in)
Weight105g approx. (0.23lb approx.)
Top of Page

NRF40

NRF40
Radiation MeasuredGamma (X) rays
DetectorSilicon semiconductor
Measurement Range1μSv to 9.999Sv
(0.1mrem to 999.9rem)
1μSv/h to 9.999Sv/h
(0.1mrem/h to 999.9rem/h) for dose measurement
Accuracy≤ ± 10%
(0.1mSv to 9.999Sv (10mrem to 999.9rem), 137Cs)
Energy Range30keV to 6MeV
Energy Response≤ ± 20% (50keV to 6MeV, 137Cs reference)
Angular Response≤ ± 20% (Up to ±60degree vertical and horizontal, 137Cs)
≤ ± 50% (Up to ±60degree vertical and horizontal, 241Am)
≤ ± 30% (All around horizontal,137Cs, free air)
Dose Rate Linearity≤ ± 10%
(0.1mSv/h to 9.999Sv/h (10mrem/h to 999.9rem/h),137Cs)
DisplayLCD (with backlight)
Alarm Volume85 to 100dB (at 30cm)
Communication MethodInfrared communication
Operating Temperature-20℃ to 50℃ (-4°F to 122°F)
Operating Humidity35% to 95%RH (Non-condensing)
Power SupplyPrimary Battery (CR123A) x 1
Battery Life2880hours (without alarm/ communication)
Dimensions62(W) x 35(D) x 82(H)mm
(2.4(W) x 1.4(D) x 3.2(H) in)
Weight115g approx. (0.25lb approx.)