MICRO+™ MR

MR-Compliant Fixed Lasers for RT & Diagnostic Imaging

For departments practicing MRgRT, MICRO+™ MR offers remote control laser positioning in an MR environment. It’s MR-compatible, up to 3 Tesla.

 

 

Accuracy at the Patient

With laser line widths of ≤0.5 mm for all colors, and industry-leading lengths of ≥4 meters at 3 meters, MICRO+ MR offers precision and flexibility for your workflow.

 

Remote Controlled, Workflow Optimized

Lose the ladder, with easy adjustments via an included remote control. Backlit for visibility in darkly lit rooms, the remote easily synchronizes and controls multiple lasers with Bluetooth technology (no line of sight needed). A locator feature helps in case of misplacement.

MR-Ready
The system mounts to walls or ceiling, keeping MR rooms organized, and includes an MRI-compliant power supply kit.


 


.

Specifications

Device Specifications

Adjustment Type

Handheld remote control

Degrees of Movement/Freedom

6

Left – Right

≥ ± 15 mm

Up – Down

≥ ± 15 mm

Rotation

≥ ± 5°

Horizontal Tilt (yaw)

≥ ± 5°

Vertical Tilt (pitch)

≥ ± 5°

Focus Range

1.5 m – 4 m

Adjustment Accuracy

0.15 mm

Adjustments Speed

Slow – Ultra accurate steps (0.15 mm) for each touch of the control

Medium – Hold the control for continuous motion

Fast – Continue to hold the control for faster speeds and larger movements

Remote Technology

Bluetooth & infrared

Remote Receiver

Integrated

Remote Operational Range

>10 m

Remote Locator

“Find remote” option on each laser

Number of Lasers per Remote

6

Laser Selection

OneTouch automatic laser and remote pairing

Line Width (All Colors)

≤ 0.5 mm for all colors @ 4 m

Line Length

≥ 4 meter @ 3 m

Laser Type

Diode

Laser Output

< 1 mW

Power Supply

100 – 240 volt (auto select)

Dimensions

216 mm × 134 mm × 80 mm (H×W×D)

Weight

< 1 kg

Available Colors

Wavelength 635 nm Red

Wavelength 515 nm Green

Wavelength 450 nm Blue

Installation Options

Wall mount ± 45° (with optional tilt and adapter bracket)

Certifications

Gammex is a ISO13485 certified medical device manufacturer. Made in the USA. Complies with Center for Devices and Radiological Health regulations for Class II lasers and all CE requirements.