SAM 950 (RIID) Ruggedized Radiation Isotope Identifier
The Model SAM 950 Ruggedized Isotope Identifier is designed to meet the requirements of ANSI 42.34 criteria for hand-held instruments for the detection and identification of radionuclides. Our patented hand-held RIID feature provides a superb solution for the demanding requirements of Homeland Security in the USA and other countries. This ruggedized handheld radiation detection system is designed specifically for first responders, field analysis of nuclear materials and challenging military applications. BNC is a world leader in the manufacturing and development of handheld radionuclide identification devices.

OVERVIEW
Quick Specs | Applications |
|
|
- Designed to meet all ANSI N42.34 requirements
- Ruggedized Design – Water & Shock Resistant
- PDA based hand-held RIID (patented)
- One click reach back connect (send files automatically)
- Automatic System Calibration and Stabilization
- Database protocol for log files and alarm events
- Solid state or 3He neutron detector
- Multimedia support
- LED flash light for target identification
- Touch screen with gesture control
- NOTE – WEB BASED TRAINING AVAILABLE NOW
- https://academy.berkeleynucleonics.com/
- Homeland Security
- Emergency Responder
- Radiological Area Mapping
- Geological Radiation Survey
- Safeguard and Nuclear Security
- Touch Screen Easy to use Reach-Back via E-mail
SPECIFICATIONS
INPUT / OUTPUT
USB
Micro USB 2.0, IP65
AC Adapter
12V, 3A
Camera
CMOS 13.0 MP
Speaker
84dB(1W/0.5m), Water Proof
GPS
Built-in GPS in Smartphone
Display
5 inch Touch Screen
PHYSICAL
Dimensions (W x D x H)
192 x 356 x 214 (mm)
Weight
4.3kg (9.5 lb) w/ 3×3 inch NaI(Tl)
3.2kg (7 lb) w/ 2×2 inch NaI(Tl)
3kg (6.6 lb) w/ 1.5×1.5 inch LaBr3 or CeBr3
ENVIRONMENTAL
Operating Temperature
-15℃(5℉)~50℃(122℉)
Relative Humidity
10 to 80%, non condensing
PERFORMANCE
Energy Resolution (Gamma)
NaI(Tl) 2×2, 3×3 inch : < 7%@662 Kev,
LaBr3 2×2 inch : < 3% @662Kev,
CeBr3 2×2 inch : < 4% @662Kev
Energy Range (Gamma)
20 keV – 3 MeV
MCA channel
10bit, 1024 channel
Dose rate range
0 – 10 mR/h (NaI), 10 mR/h – 10R/h (GM)
Stabilization
Automatic real-time stabilization using K-40
Nuclide Identification
According to ANSI N42.34, isotope/category/confidence report
Battery
> 8 hours, Lithium Ion